Mostrar el registro sencillo del ítem

dc.contributor.authorLozano Martín, Daniel 
dc.contributor.authorMondéjar Montagud, María Engracia
dc.contributor.authorSegovia Puras, José Juan 
dc.contributor.authorChamorro Camazón, César Rubén 
dc.date.accessioned2019-10-31T09:14:07Z
dc.date.available2019-10-31T09:14:07Z
dc.date.issued2019
dc.identifier.citationDaniel Lozano-Martín, María E. Mondéjar, José J. Segovia, César R. Chamorro. Determination of the force transmission error in a single-sinker magnetic suspension densimeter due to the fluid-specific effect and its correction for use with gas mixtures containing oxygen. Measurement (2019). DOI: https://doi.org/10.1016/j.measurement.2019.107176es
dc.identifier.issn0263-2241es
dc.identifier.urihttp://uvadoc.uva.es/handle/10324/38991
dc.descriptionProducción Científicaes
dc.description.abstractDensity measurements from single-sinker magnetic suspension densimeters need to be corrected to compensate for the magnetic effects of the measuring cell materials and the fluid on the coupling transmission system. While the magnetic effect of the densimeter materials can be easily determined, the fluid effect requires the calculation of an apparatus-specific constant, ερ. In this work, the apparatus-specific constant of the single-sinker magnetic suspension densimeter at the University of Valladolid has been determined by using two alternative methods. The first method, which uses density data for the same fluid and conditions and different sinkers, yielded a value of ερ = 4.6·10-5. The second method, obtained from measurements with pure oxygen, yielded a value of ερ = 8.822·10-5. The second value is considered as more reliable, as the first method presents inherent limitations in this case.es
dc.format.mimetypeapplication/pdfes
dc.language.isoenges
dc.publisherElsevieres
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.titleDetermination of the force transmission error in a single-sinker magnetic suspension densimeter due to the fluid-specific effect and its correction for use with gas mixtures containing oxygenes
dc.typeinfo:eu-repo/semantics/articlees
dc.rights.holderElsevieres
dc.identifier.doi10.1016/j.measurement.2019.107176es
dc.relation.publisherversionhttps://doi.org/10.1016/j.measurement.2019.107176es
dc.identifier.publicationfirstpage107176es
dc.identifier.publicationtitleMeasurementes
dc.peerreviewedSIes
dc.description.projectMinisterio de Economía, Industria y Competitividad project ENE2017-88474-Res
dc.description.projectJunta de Castilla y León project VA280P18es
dc.type.hasVersioninfo:eu-repo/semantics/acceptedVersiones