Mostrar el registro sencillo del ítem

dc.contributor.authorValencia-Nieto, Laura
dc.contributor.authorLópez-de la Rosa, Alberto
dc.contributor.authorGonzález García, María J.
dc.contributor.authorLópez-Miguel, Alberto
dc.date.accessioned2024-08-30T08:59:18Z
dc.date.available2024-08-30T08:59:18Z
dc.date.issued2024
dc.identifier.citationOphthalmic and Physiological Optics, Enero 2024, vol. 44, n. 1. p. 124-130.es
dc.identifier.issn0275-5408es
dc.identifier.urihttps://uvadoc.uva.es/handle/10324/69560
dc.descriptionProducción Científicaes
dc.description.abstractPurpose: To assess the reliability and agreement of non-invasive break-up time (NIBUT) in symptomatic and asymptomatic contact lens (CL) wearers using automatic objective and conventional subjective techniques. Methods: In this prospective cross-sectional study, soft CL wearers, classified into symptomatic and asymptomatic based on the Contact Lens Dry Eye Questionnaire-8, underwent NIBUT assessment with the CL in situ. The CA-800 Corneal Analyzer and the EasyTear® VIEW+ Tearscope were used for objective and subjective evaluation, respectively. The within-subject repeatability and intraclass correlation coefficient (ICC) were calculated. The agreement between the devices was compared using the Bland–Altman method. Results: A total of 141 CL wearers (51 male and 90 female) with a mean age of 33.6 (SD = 12.2) years were included. The repeatability and ICC values obtained with the CA-800 device when measuring NIBUT were 5.4 s and 58.6% across the whole sample, 4.2 s and 48.8% for the asymptomatic group and 7.1 s and 68.4% for the symptomatic group. When using the subjective method (EasyTear®), the respective repeatability and ICC values were 7.3 s and 32.7% for the whole sample, 6.5 s and 30.4% for the asymptomatic group and 8.6 s and 35.9% for the symptomatic group. The CA-800 device provided significantly (p < 0.001) shorter NIBUT values compared with EasyTear® for the whole sample (3.3 [2.9] vs. 8.1 [3.4] s), the asymptomatic (3.3 [3.0] vs. 7.7 [3.6] s) and the symptomatic (3.8 [2.9] vs. 8.6 [3.0] s) groups. Conclusion: Objective (CA-800) NIBUT assessment provides more reliable measurements than the conventional subjective technique using the EasyTear® device. However, CL practitioners should also be aware that the objective method indicates shorter NIBUT values. Symptomatic CL wearers may also need a higher number of NIBUT measurements to obtain reliable estimations.es
dc.format.mimetypeapplication/pdfes
dc.language.isoenges
dc.publisherWileyes
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectLentes de contactoes
dc.subject.classificationAgreementes
dc.subject.classificationCA-800es
dc.subject.classificationContact lenses
dc.subject.classificationEasyTear® VIEW+es
dc.subject.classificationNon-invasive break-up timees
dc.subject.classificationRepeatabilityes
dc.titleReliability and agreement of subjective and objective non-invasive break-up time measurements in contact lens wearerses
dc.typeinfo:eu-repo/semantics/articlees
dc.identifier.doihttps://doi.org/10.1111/opo.13243es
dc.peerreviewedSIes
dc.description.projectMinistry of Universities and European Social Fund, Grant/Award Number: FPU19/01109es
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.type.hasVersioninfo:eu-repo/semantics/publishedVersiones
dc.subject.unesco3201.09 Oftalmologíaes


Ficheros en el ítem

Thumbnail

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem