RT info:eu-repo/semantics/article T1 A measurement setup and automated calculation method to determine the charge injection capacity of implantable microelectrodes A1 Fraile Marinero, Juan Carlos A1 Pérez Turiel, Javier A1 Muñoz Martinez, Victor A1 Müller, Carsten A1 Ihmig, Frank K1 Access voltage K1 Cyclic voltammetry K1 Voltage transient measurements K1 Charge injection capacit K1 Charge storage capacity K1 Surface roughness K1 33 Ciencias Tecnológicas AB The design of safe stimulation protocols for functional electrostimulation requires knowledge of the “maximum reversible charge injection capacity” of the implantable microelectrodes. One of the main difficulties encountered in characterizing such microelectrodes is the calculation of the access voltage Va. This paper proposes a method to calculate Va that does not require prior knowledge of the overpotential terms and of the electrolyte (or excitable tissue) resistance, which is an advantage for in vivo electrochemical characterization of microelectrodes. To validate this method, we compare the calculated results with those obtained from conventional methods for characterizing three flexible platinum microelectrodes by cyclic voltammetry and voltage transient measurements. This paper presents the experimental setup, the required instrumentation, and the signal processing. PB MDPI SN 1424-8220 YR 2018 FD 2018 LK https://uvadoc.uva.es/handle/10324/52531 UL https://uvadoc.uva.es/handle/10324/52531 LA eng NO Sensors, 2018, vol. 18, n. 12, p. 4152 NO Producción Científica DS UVaDOC RD 26-abr-2024