RT info:eu-repo/semantics/article T1 Reliability and agreement of apparent chord mu measurements between static and dynamic evaluations A1 Martínez Plaza, Elena A1 Marcos, Mario A1 López De La Rosa, Alberto A1 López Miguel, Alberto A1 Maldonado López, Miguel José K1 Chord mu K1 Scheimpflug and Placido-disk technology K1 repeatability K1 agreement AB Purpose: To assess the repeatability and agreement of Cartesian coordinates and the length of apparent chord mu and pupil diameter measurements during static (Galilei G4) and dynamic (Topolyzer Vario) evaluations.Setting: IOBA-Eye Institute, Valladolid, Spain.Design: Case series.Methods: 3 consecutive measurements per scenario (Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions) were performed by the same clinician. The intrasession repeatability was assessed using the within-subject SD (Sw), the precision, the coefficient of variation, and the intraclass correlation coefficient (ICC). The agreement was analyzed using repeated-measures analysis of variance and the Bland-Altman method.Results: Thirty-seven healthy participants were recruited. The Sw values for chord mu parameters and pupil diameter ranged from 0.01 to 0.03 and 0.08 to 0.21, respectively. The ICC was ≥0.89 for all parameters. Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions provided significantly different measures of apparent chord mu length (0.23 ± 0.11 mm, 0.30 ± 0.10 mm, and 0.25 ± 0.11 mm, respectively, P ≤ .02), X-coordinate (-0.18 ± 0.12 mm, -0.27 ± 0.11 mm, and -0.21 ± 0.12 mm, respectively, P < .001), and pupil diameter (3.38 ± 0.50 mm, 6.29 ± 0.60 mm, and 3.04 ± 0.41 mm, respectively, P < .001). Y-coordinate values obtained by Galilei G4 and Topolyzer Vario under low mesopic conditions were significantly different (0.06 ± 0.13 mm vs 0.03 ± 0.11 mm, respectively, P = .02), in contrast to Galilei G4 and Topolyzer Vario under photopic conditions (0.05 ± 0.13 mm, P = .82) and both illumination conditions of Topolyzer Vario (P ≥ .23).Conclusions: Galilei G4 and Topolyzer Vario provide consistent measurements of apparent chord mu Cartesian coordinates and length, as well as pupil diameter; however, the measurements are not interchangeable. Ophthalmic surgeons should consider these findings when planning customized intraocular lens implantation and refractive surgery procedures. SN 0886-3350 YR 2023 FD 2023 LK https://uvadoc.uva.es/handle/10324/75174 UL https://uvadoc.uva.es/handle/10324/75174 LA eng NO Journal of Cataract & Refractive Surgery, 2023, vol. 49, n. 1, p. 21-28. DS UVaDOC RD 09-abr-2025