TY - JOUR AU - Lozano Martín, Daniel AU - Mondéjar, María E. AU - Segovia Puras, José Juan AU - Chamorro Camazón, César Rubén PY - 2020 SN - 0263-2241 UR - http://uvadoc.uva.es/handle/10324/43191 AB - Density measurements from single-sinker magnetic suspension densimeters need to be corrected to compensate for the magnetic effects of the measuring cell materials and the fluid on the coupling transmission system. While the magnetic effect of the... LA - eng PB - Elsevier KW - Termodinámica KW - force transmission error; magnetic-suspension coupling; fluid-specific effect; apparatus-specific constant; single sinker densimeter TI - Determination of the force transmission error in a single-sinker magnetic suspension densimeter due to the fluid-specific effect and its correction for use with gas mixtures containing oxygen DO - 10.1016/j.measurement.2019.107176 ER -