Safety, biocompatibility, and potential functionality of a new accommodative intraocular lens: An experimental study in rabbits
Autor
Año del Documento
2023
Editorial
Elsevier
Descripción
Producción Científica
Documento Fuente
Heliyon, 2023, vol. 9, issue 9, e19604
Résumé
This study aims to evaluate the safety, biocompatibility, and functionality of a new accommodative intraocular lens (IOL) (LUZ, patent PCT/ES2016/070,813) after implantation in rabbit eyes. LUZ (Study) and EyeCee® plus a capsular ring (Control) were implanted in rabbits (n = 8 each) after phacoemulsification. Intraoperative follow-up, long-term clinical follow-up, and functional IOL studies were carried out periodically for up to 180 days. A macroscopic examination of the eyeballs to reveal abnormalities and determine the implant centering and a microscopic examination to semi-quantify cell and tissue response were performed. Statistical analysis of the collected data was finally achieved. During follow-up, no significant changes in the general condition nor the clinical evaluation were observed between both groups. However, Study IOL remained centered throughout the study and did not present severe complications as observed in the Control group. Functional studies did not reveal significant differences between both materials. Study showed better centering, fewer adhesions, and maintenance of an opening capsular bag compared to the Control. Local biological effects caused by Study implantation are minimal and comparable to the Control. Therefore, LUZ showed no clinical signs or histological response of adverse reaction to the implanted material, according to UNE-EN ISO 11979-5 and 10993-6. Functionality must be confirmed in another animal species with greater lens accommodation capacity than the rabbit. LUZ keeps the capsular bag open, favoring its centering and avoiding fibrosis and adherence to the bag; this allows potential accommodation of this IOL and theoretically enables the patient to focus dynamically.
Materias (normalizadas)
Eye Diseases
Oftalmología
Materias Unesco
3201.09 Oftalmología
Palabras Clave
Presbyopia
Intraocular lens
Biocompatibility
Presbicia
Lente intraocular
Biocompatibilidad
ISSN
2405-8440
Revisión por pares
SI
Patrocinador
Ministerio de Ciencia e Innovación (Neotec Project EXP-00104711/SNEO-20171172)
Propietario de los Derechos
© 2023 The Authors
Idioma
eng
Tipo de versión
info:eu-repo/semantics/publishedVersion
Derechos
openAccess
Aparece en las colecciones
Fichier(s) constituant ce document
Excepté là où spécifié autrement, la license de ce document est décrite en tant que Attribution-NonCommercial-NoDerivatives 4.0 Internacional