Por favor, use este identificador para citar o enlazar este ítem:https://uvadoc.uva.es/handle/10324/73293
Título
Automated analysis of 2-methyl-3-furanthiol and 3-mercaptohexyl acetate at ngL−1 level by headspace solid-phase microextracion with on-fibre derivatisation and gas chromatography–negative chemical ionization mass spectrometric determination
Año del Documento
2006-07-14
Editorial
Elsevier
Documento Fuente
Journal of Chromatography A, Volume 1121, Issue 1, 14 July 2006, Pages 1-9
Résumé
A fast and automated method for the analysis at ng L−1 level of aroma-powerful polyfunctional thiols has been developed and applied to wine. The sample is just poured in a 20 mL vial and its vapour extracted with a poly(dimethylsiloxane)–divinylbenzene (PDMS–DVB) solid-phase microextraction fibre (65 μm thickness) previously exposed to vapours of the reactive (pentafluorobenzyl bromide) and of an alkali (tributylamine). The derivatised compounds are subsequently desorbed in the GC system and determined by negative chemical ionization mass spectrometry. The method is fully automated by using a Combi-Pal autosampler conveniently programmed. The analysis takes 50 min, which contrasts to the long and tedious methods previously proposed. The development of an optimal procedure is constrained by the aggressive character of the reagent (towards the fibre and the chromatographic column), its volatility and the quality of the blanks that can be obtained. Therefore, a critical step was fixing in the fibre a “safe” and repetitive amount of reagent. This was achieved by exposing the fibre (5 min) to the vapours of a water:acetone (9:1) solution containing 200 mg L−1 of reagent. Under these conditions, the extraction-derivatisation of analytes improves with time and temperature, and the best working conditions are dictated by a compromise between sensitivity, speed and chromatographic performance. Although analytes studied were 2-methyl-3-furanthiol, 4-mercapto-4-methyl-2-pentanone, 3-mercaptohexanol, 2-furanmethanethiol and 3-mercaptohexyl acetate, a good analytical performance could be achieved only for these two last compounds. Both of them can be repetitively (10% < RSD < 20%) determined in wine at concentrations below 0.1 ng L−1. Other aspects considered in the method setup were the oxidation of analytes during the process, and the electron-capture detection (ECD) and MS properties of the pentafluorobenzyl derivatives of different polyfunctional thiols.
Palabras Clave
GC–MS–NCI
Thiols
SPME
ISSN
0021-9673
Revisión por pares
SI
Propietario de los Derechos
Elsevier
Idioma
spa
Tipo de versión
info:eu-repo/semantics/publishedVersion
Derechos
embargoedAccess
Aparece en las colecciones
Fichier(s) constituant ce document
Nombre:
1-s2.0-S0021967306007539-main.pdfEmbargado hasta: 2026-01-09
Tamaño:
340.0Ko
Formato:
Adobe PDF