Intrasession repeatability and agreement of a new method to measure the foveal fixation axis
Año del Documento
2023
Descripción
Producción Científica
Documento Fuente
PeerJ, Febrero 2023, vol. 24, n. 11. p. e14942
Zusammenfassung
Purpose: Ophthalmic lens adaptation, particularly with progressive addition lenses, requires accurate measurements of the patient nasopupillary distance (NPD) and interpupillary distance (IPD), which are usually collected using the pupil centre as a reference. However, differences between the pupil centre and visual or foveal axis could induce some subsidiary effects of correcting lenses. This study aimed to assess the intrasession repeatability of a new prototype (Ergofocus®; Lentitech, Barakaldo, Spain) that can measure the foveal fixation axis (FFA) distance and assess the agreement with the NPD measurements collected using a traditional method (frame ruler).
Methods: The FFA at far and near distances was measured three consecutive times in 39 healthy volunteers to determine the intrasession repeatability according to the British Standards Institute and International Organization for Standardization. Additionally, the FFA and NPD (standard frame ruler) were measured in 71 healthy volunteers and compared using Bland-Altman analysis. Two blinded experienced practitioners conducted each FFA and NPD measurement.
Results: The FFA measurements showed acceptable repeatability at far distances (right eye (RE): Sw = 1.16 ± 0.76 mm and coefficient of variation (CV) = 3.92 ± 2.51%; left eye (LE) Sw = 1.11 ± 0.79 mm and CV = 3.76 ± 2.51%) and at near distances (RE: Sw = 0.97 ± 0.85 mm and CV = 3.52 ± 3.02%; LE: Sw = 1.17 ± 0.96 mm and CV = 4.54 ± 3.72%). Additionally, agreement with the NPD showed large differences at far distances (RE: -2.15 ± 2.34, LoA = -6.73 to 2.43 mm (P < 0.001); LE: -0.61 ± 2.62, LoA = -5.75 to 4.53 mm (P = 0.052)) and near distances (RE: -3.08 ± 2.80, LoA -8.57 to 2.42 mm (P < 0.001); LE: -2.97 ± 3.97, LoA: -10.75 to 4.80 mm (P < 0.001)).
Conclusions: FFA measurements showed clinically acceptable repeatability at both far and near distances. Agreement with the NPD measured using a standard frame ruler showed significant differences, suggesting that both measurements are not interchangeable in clinical practice to prescribe and center ophthalmic lenses. Further research is necessary to assess the impact of FFA measurement in ophthalmic lens prescriptions.
Materias (normalizadas)
Optometría
Palabras Clave
Agreement
Foveal fixation axis
Intrasession repeatability
Nasopupillary distance
Ophthalmic lens prescription